V-Go

Basal delivery,
Continuous subcutaneous
20 Units/24 hr (0.83 U/hr)
30 Units/24 hr (1.25 U/hr)
40 Units/24 hr (1.67 U/hr)
Bolus delivery,
on demand
Each click of the Bolus Delivery Button provides 2 Units
Each V-Go contains 18 uses for a maximum of 36 Units per V-Go
V-Go minimum capacity reservoir 0.56 mL V-Go 20
0.66 mL V-Go 30
0.76 mL V-Go 40
V-Go size 2.4 x 1.3 x 0.5 inches
(6.1 cm x 3.3 cm x 1.3 cm)
V-Go weight 0.7 to 1.8 ounces
(20 to 50 grams)
Storage conditions— device only +10°F (-12°C) to +140°F (+60°C)
20% to 90% relative humidity
Avoid prolonged exposure to the extremes of the storage temperature range
Storage conditions— filled with Humalog® V-Go may be filled with Humalog for up to 24 hours prior to use if refrigerated or if left at room temperature
Storage conditions— filled with NovoLog® V-Go may be filled with NovoLog for up to 5 days prior to use if refrigerated. V-Go may be filled with NovoLog for up to 3 days prior to use if left at room temperature
Operating conditions +40°F (+5°C) to +99°F (+37°C)
20% to 90% relative humidity
Do not expose V-Go to temperatures over 99°F for prolonged periods of time (beyond 14 hours) as this may lead to degradation of the insulin (check the insulin manufacturer’s instructions for use for temperature details) and may cause V-Go to run at a basal rate slightly greater than the specified range
Operating atmospheric pressure 697 hPa to 1,013 hPa
(Equivalent to altitudes up to 10,000 feet [3,048 meters])
Duration of use 24 hours
Bolus delivery accuracy (for all 3 V-Go options) +/-10%
Basal delivery accuracy (for all 3 V-Go options)  +/-10%
V-Go power source Mechanical
Enclosure protection  Protects against the effects of submersion at depths up to 3 feet, 3 inches (1 meter) for 24 hours. (IPX8)
Needle 4.6-mm, 30-gauge needle with sharps injury prevention features and floating needle technology
Disposal Local disposal requirements

EZ Fill

Storage conditions— unused device only -4°F (-20°C) to +140°F (+60°C)
20% to 90% relative humidity
Operating conditions +40°F (+5°C) to +99°F (+37°C)
20% to 90% relative humidity
Warm to room temperature to optimize fill quality
Insulin Each time before placing a new vial into the EZ Fill, check to make sure the insulin vial is not expired and that it will not expire during the time period that the vial will be used in the EZ Fill
Duration of use 30 fills
Disposal  Local disposal requirements